Model
DH-21
Owner's Manual
NOTICE
This product embodies technology that is confidential and proprietary technology of DESIGN
ANALYSIS ASSOCIATES, INC., and which is protected by United States copyright laws and
international copyright treaty provisions, and/or by contract and applicable laws of trade secrecy.
These include all Software, Printed Circuit Board Artwork, Schematic Diagrams, and Technologies
applied therein. The enclosure encasing the electronics of this instrument may not be opened
without written consent of DESIGN ANALYSIS ASSOCIATES, INC., and any attempt to do so
without such written authorization constitutes a breach of contract and will also void any applicable
warranty for the product.
Document Number: DH21001, DH21002
Document Authors: Terrell Fletcher
Revision: 1.5
75 West 100 South
Logan, UT 84321 USA
Phone: (435) 753-2212
Fax: (435) 753-7669
Zobrazit stránku 0
1 2 3 4 5 6 ... 111 112

Shrnutí obsahu

Strany 1 - Owner's Manual

ModelDH-21Owner's ManualNOTICEThis product embodies technology that is confidential and proprietary technology of DESIGNANALYSIS ASSOCIATES, INC.

Strany 2

1-2 Introduction DH-21Figure 1-1. The “WATERLOG” Sensor/Logger1.2 Unpacking the “WATERLOG” Here are the items you should have received:1. WATERLOG® S

Strany 3 - Table of Contents

4-54 Using the Graphical “WATERLOG” Software DH-21Figure 4-37. Self-Test Diagnostic Display 4.4.12 File OptionThe File Options shown in Figure 4-38

Strany 4

DH-21 Using the Graphical “WATERLOG” Software 4-55The options that are saved are listed below.Logging modes.Conversion optionsUser notesView graph op

Strany 5 - Warranty

4-56 Using the Graphical “WATERLOG” Software DH-21Figure 4-38. File Options 4.4.13 Attach OptionThe Attach option shown in Figure 4-39 allows the “W

Strany 6

DH-21 Using the Graphical “WATERLOG” Software 4-57In the stand-alone mode the “WATERLOG” software will not display all status information of thesenso

Strany 7

4-58 Using the Graphical “WATERLOG” Software DH-214.4.14 Help OptionGeneral Help shown in Figure 4-40 gives a basic understanding about the differen

Strany 8

Chapter 5Guide to Data AnalysisDH-21 Guide to Data Analysis 5-1Data is placed in a file on your DOS disk under the filename you specified with a “.DA

Strany 9 - Introduction

5-2 Guide to Data Analysis DH-21“WaterLOG Logging Module (All data extracted)”“Serial Number : T2000"“Extracted : 08:11:18 03/02/95"“L

Strany 10 - WATERWARE

Chapter 6Maintenance/Trouble ShootingDH-21 Maintenance/Trouble Shooting 6-16.1 MaintenanceThe “WATERLOG” was designed to be a low maintenance instrum

Strany 11 - DH-21 Introduction 1-3

6-2 Maintenance/Trouble Shooting DH-21No Response 1) Verify power to the “WATERLOG”.2) Make sure you have connected the RS-232 cable properly.3) Make

Strany 12

Appendix ASpecificationsDH-21 Specifications A-1SENSOR SPECIFICATIONSAccuracy(Maximum percent of error in measurement)Pressure: For measurement inter

Strany 13 - Installation

DH-21 Introduction 1-31.3 Checking the Model NumberBefore installing your new “WATERLOG”, check the information on the label of the sensorenclosure.

Strany 14 - WATERLOG” is

A-2 Specifications DH-21Environmental RestrictionsOperating Range: 0 to 40C (nonfreezing)Compensated Range: 0 to 40CStorage: -10 to 70 C Extende

Strany 15 - DH-21 Installation 2-3

DH-21 Specifications A-3GENERAL SPECIFICATIONSPower SupplyPrimary Source: 2 Lithium cells at 3.6 V, 13.5 AH each (User Replaceable)Or Alternate Exter

Strany 18

Chapter 2InstallationDH-21 Installation 2-12.1 Installing the “WATERLOG”This chapter covers installation techniques for the “WATERLOG”. For proper in

Strany 19 - Using the “WATERLOG” Software

2-2 Installation DH-21 Figure 2-1. Typical Small Bore Well Installation2.1.2 Sample Well InstallationThe “WATERLOG” can be used inside a w

Strany 20

DH-21 Installation 2-32.1.3 Environmental ConcernsThrough years of experience dealing with measuring in the environment, we have learned thatMother N

Strany 21 - WATERLOG”. The

2-4 Installation DH-21Figure 2-2. RS-232 Cable2.3 Accessing the Battery Pack and DesiccantThe “WATERLOG” uses a factory assembled 7.2-Volt, 13.5 Amp H

Strany 22 - WATERLOG”

DH-21 Installation 2-5 Figure 2-3 Accessing the Battery Pack and DesiccantDesiccant Cartridge can be reused by drying them in an

Strany 24

Chapter 3Using the “WATERLOG” SoftwareDH-21 Using the “WATERLOG” Software 3-1Note: The communications cable to the “WATERLOG” Logger must be connec

Strany 26 - ATERLOG” Software DH-21

3-2 Using the “WATERLOG” Software DH-21The screen of the computer is updated every 5 seconds with the logger time, date and data. When the WATERGEN p

Strany 27 - ATERLOG” Software 3-9

DH-21 Using the “WATERLOG” Software 3-3The Memory Used field indicates what percentage of the FLASH EEPROM currently has beenfilled. The FLASH EEPR

Strany 28

3-4 Using the “WATERLOG” Software DH-21Warning : This option should not be invoked if the user has uncollected data in theFLASH EEPROM which needs t

Strany 29 - WATERLOG's” internal

DH-21 Using the “WATERLOG” Software 3-53.2.2 Time/Date OptionYou can invoke menu item 4 as shown in Figure 3-3. The correct date and time must be t

Strany 30

3-6 Using the “WATERLOG” Software DH-213.2.3 Logging Mode OptionOptions 5, 6, and 7 configure the “WATERLOG” for unattended logging. Option 5 is sho

Strany 31 - WATERLOG” will be

DH-21 Using the “WATERLOG” Software 3-7There are five different sampling modes the “WATERLOG” sensor can operate under. EachLogging Mode is briefly

Strany 32

3-8 Using the “WATERLOG” Software DH-213.2.3.1 Linear Logging Mode OptionThe Linear Logging Mode shown in Figure 3-5 is configured to have a constant

Strany 33 - WATERLOG” sensor

DH-21 Using the “WATERLOG” Software 3-9This option is handy for saving memory space since data is not logged if the data does not changemore than th

Strany 34

3-10 Using the “WATERLOG” Software DH-21Hyper Divisor: The Hyper Divisor is used to determine the faster scan rate. The faster scan rateis calculate

Strany 35

DH-21 Using the “WATERLOG” Software 3-11 The WaterLOG Serial # T2000 Version 1.1 

Strany 36

Table of ContentsUser Agreement/WATERLOG® Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . W-1Chapter 1 Intr

Strany 37 - ATERLOG” Software 3-19

3-12 Using the “WATERLOG” Software DH-21is calculated by taking the regular scan rate and dividing it by the hyper divisor. The hyper divisor must be

Strany 38 - WATERLOG” to

DH-21 Using the “WATERLOG” Software 3-13 The WaterLOG

Strany 39 - ATERLOG” Software 3-21

3-14 Using the “WATERLOG” Software DH-21As mentioned, this mode is used to simulate a logarithmic logging mode. Below is a simplesetup showing a log

Strany 40 - WATERLOG” must be

DH-21 Using the “WATERLOG” Software 3-15Note: It is important to program these values correctly. The data displayed in the last measurementwindow

Strany 41 - WATERLOG” for the

3-16 Using the “WATERLOG” Software DH-21The “WATERLOG” stores 160 characters of User Notes in the FLASH EEPROM. These notesallow you to enter inform

Strany 42

DH-21 Using the “WATERLOG” Software 3-173.2.6 Run OptionThere are three basic modes of operation of the “WATERLOG”. Up to now the “WATERLOG”has bee

Strany 43 - ATERLOG” Software 3-25

3-18 Using the “WATERLOG” Software DH-21 The WaterLOG

Strany 44

DH-21 Using the “WATERLOG” Software 3-19 The WaterLOG

Strany 45 - ATERLOG” Software 3-27

3-20 Using the “WATERLOG” Software DH-21Shown in Figure 3-16, the software running in the computer will cause the “WATERLOG” totransfer the data to t

Strany 46

DH-21 Using the “WATERLOG” Software 3-21 The WaterLOG

Strany 47 - Chapter 4

Chapter 4 Using the Graphical “WATERLOG” Software 4.1 Software Overview and Installation . . . . . . . . . . . . .

Strany 48 - WATERLOG” will change to a

3-22 Using the “WATERLOG” Software DH-21Warning: If the power has failed in the “WATERLOG”, the FLASH EEPROM will still containthe data sampled up t

Strany 49 - ATERLOG” Software 4-3

DH-21 Using the “WATERLOG” Software 3-233.2.10 Measure Data OptionThe Measure Data Option causes the “WATERLOG” to take a measurement as shown in F

Strany 50

3-24 Using the “WATERLOG” Software DH-21 The WaterLOG

Strany 51 - ATERLOG” Software 4-5

DH-21 Using the “WATERLOG” Software 3-25 The WaterLOG

Strany 52

3-26 Using the “WATERLOG” Software DH-21 The WaterLOG

Strany 53 - ATERLOG” Software 4-7

DH-21 Using the “WATERLOG” Software 3-27 Wat

Strany 55 - ATERLOG” Software 4-9

Chapter 4Using the Graphical "WATERLOG" SoftwareDH-21 Using the Graphical “WATERLOG” Software 4-14.1 Software Overview and Installation Thi

Strany 56 - WATERLOG” will point to

4-2 Using the Graphical “WATERLOG” Software DH-21The following will be displayed on the screen.Wrong number of parameters have been specified.WATERL

Strany 57

DH-21 Using the Graphical “WATERLOG” Software 4-3Figure 4-1. Master Menu ScreenFigure 4-1 shows the master menu screen. This screen is divided into s

Strany 58 - Figure 4-5. Run Option

User Agreement/WATERLOG® WarrantyDH-21 User Agreement/WATERLOG® Warranty W-11. NATURE OF THE PRODUCTThis agreement accompanies a pressure measuring

Strany 59 - ATERLOG” Software 4-13

4-4 Using the Graphical “WATERLOG” Software DH-21the actual percent full. As the memory fills, warning messages will be displayed at thebottom of th

Strany 60

DH-21 Using the Graphical “WATERLOG” Software 4-5Line 5, Logging Rate:The logging rate is a time from 1 second to 24 hours in the form of 00:00:01 to

Strany 61

4-6 Using the Graphical “WATERLOG” Software DH-21Main Options Window:The main option buttons allow full control of the sensor when it is connected,

Strany 62

DH-21 Using the Graphical “WATERLOG” Software 4-7will cause the sensor to record levels with a -5.0 foot offset. The logged value isbased on the foll

Strany 63 - ATERLOG” Software 4-17

4-8 Using the Graphical “WATERLOG” Software DH-21Box in upper right corner:When there is a small box in the upper right corner, it is used to maximi

Strany 64

DH-21 Using the Graphical “WATERLOG” Software 4-9Figure 4-2. Reset OptionFigure 4-3 shows the same screen as Figure 4-2 except the Operation Mode box

Strany 65 - ATERLOG” Software 4-19

4-10 Using the Graphical “WATERLOG” Software DH-21Warning : This option should not be invoked if there is data in the FLASH EEPROM which needs tobe

Strany 66

DH-21 Using the Graphical “WATERLOG” Software 4-114.4.1.2 Stop Logging OptionThe sensor can be paused and the logging operation halted by entering th

Strany 67 - ATERLOG” Software 4-21

4-12 Using the Graphical “WATERLOG” Software DH-21Figure 4-5. Run Option In the Run Option, a start time can be specified. This will allow you to ma

Strany 68

DH-21 Using the Graphical “WATERLOG” Software 4-13Figure 4-6. Run Option/ Start TimeWith the “WATERLOG” in the RUN mode, you can disconnect the inter

Strany 69 - ATERLOG” Software 4-23

W-2 User Agreement/WATERLOG® Warranty DH-214. TERMUSER may enjoy these rights only as long as their possession of the PRODUCT shall continue tobe r

Strany 70

4-14 Using the Graphical “WATERLOG” Software DH-214.4.2 Time/Date OptionChanging the time and date is shown in Figure 4-7. The correct date and time

Strany 71 - ATERLOG” Software 4-25

DH-21 Using the Graphical “WATERLOG” Software 4-154.4.3 Logging Mode OptionLogging Mode, Coefficients, and User Notes configure the “WATERLOG” for an

Strany 72

4-16 Using the Graphical “WATERLOG” Software DH-21the new “Next Log Time” has already occurred, the “WATERLOG” will continue to add thelogging rate

Strany 73 - ATERLOG” Software 4-27

DH-21 Using the Graphical “WATERLOG” Software 4-17Figure 4-9. Linear Mode Option4.4.3.2 Delta Logging Mode OptionThe Delta Logging Mode shown in Figu

Strany 74

4-18 Using the Graphical “WATERLOG” Software DH-21This option is handy for saving memory space since data is not logged if the data does not changem

Strany 75 - ATERLOG” Software 4-29

DH-21 Using the Graphical “WATERLOG” Software 4-19Hyper Limit:The Hyper Limit will be expressed in the unit of measure specified under the conversion

Strany 76

4-20 Using the Graphical “WATERLOG” Software DH-21Figure 4-11. Hyper Mode OptionHere is an example of the Hyper Logging Mode.The scan rate is set to

Strany 77 - ATERLOG” Software 4-31

DH-21 Using the Graphical “WATERLOG” Software 4-21It is important to realize that the faster scan rate will not start until the sensor sees the chang

Strany 78

4-22 Using the Graphical “WATERLOG” Software DH-21 SCAN RATE HYPER DIVISOR FAST SCAN RATE1 minute 60 1 seconds1 minute 30 2 seconds1 minute 2

Strany 79

DH-21 Using the Graphical “WATERLOG” Software 4-23Figure 4-12. Hyper-Delta Mode Option4.4.3.5 Variable Logging Mode OptionThe Variable Logging Mode s

Strany 80

DH-21 User Agreement/WATERLOG® Warranty W-3(iii) EXCEPT FOR THE EXPRESS WARRANTIES ABOVE, DESIGNANALYSIS SPECIFICALLY DISCLAIMS ALL OTHER WARRANTIES

Strany 81 - ATERLOG” Software 4-35

4-24 Using the Graphical “WATERLOG” Software DH-21Number of Samples:The bottom half of the screen has 5 boxes labeled “Time 1” to “Time 5”. This rep

Strany 82

DH-21 Using the Graphical “WATERLOG” Software 4-25Figure 4-13. Variable Logging Mode Option4.4.4 Coefficients OptionThe Coefficients options shown in

Strany 83 - WATERLOG” must be restored

4-26 Using the Graphical “WATERLOG” Software DH-21Note: It is important to program these values correctly. The data displayed in the last measurem

Strany 84

DH-21 Using the Graphical “WATERLOG” Software 4-274.4.4.1 Pressure Coefficients OptionThe User Defined Pressure Coefficients Option shown in Figure 4

Strany 85 - ATERLOG” Software 4-39

4-28 Using the Graphical “WATERLOG” Software DH-21Figure 4-15. User Defined Pressure Coefficients Option4.4.4.2 Temperature Coefficients OptionThe U

Strany 86

DH-21 Using the Graphical “WATERLOG” Software 4-29Slope:This is the M factor of the standard Y = MX + B equation. It represents the rate of changefor

Strany 87 - WATERLOG Option

4-30 Using the Graphical “WATERLOG” Software DH-21Figure 4-16. User Defined Temperature Coefficients Option4.4.4.3 Pressure Compensation OptionThe U

Strany 88

DH-21 Using the Graphical “WATERLOG” Software 4-31Offset:The fourth option is to enter any desired offset. This may be used to add an offset to maket

Strany 89 - ATERLOG” Software 4-43

4-32 Using the Graphical “WATERLOG” Software DH-21characters each. There is no wrapping feature between lines, therefore, each line needs to beedite

Strany 90

DH-21 Using the Graphical “WATERLOG” Software 4-33Figure 4-19. Extract Data OptionShown in Figure 4-20, the software running in the computer will cau

Strany 92

4-34 Using the Graphical “WATERLOG” Software DH-21DIR:The third box down is a status box showing the current path and directory. This boxcannot be e

Strany 93 - ATERLOG” Software 4-47

DH-21 Using the Graphical “WATERLOG” Software 4-35Figure 4-21 shows a sample screen which is displayed during the Extract All Data process. Thisproce

Strany 94 - WATERLOG” software program

4-36 Using the Graphical “WATERLOG” Software DH-21the “WATERLOG” software will adjust the number to the total number of scans and startthe extract p

Strany 95 - ATERLOG” Software 4-49

DH-21 Using the Graphical “WATERLOG” Software 4-37Warning: If the power has failed in the “WATERLOG”, the FLASH EEPROM will still contain thedata sa

Strany 96

4-38 Using the Graphical “WATERLOG” Software DH-21Selecting tabular will display the data in a tabular report format. You can scroll throughthe data

Strany 97 - ATERLOG” Software 4-51

DH-21 Using the Graphical “WATERLOG” Software 4-39Figure 4-24. View Data OptionYou can examine the data stored in the FLASH EEPROM or a data file, by

Strany 98

4-40 Using the Graphical “WATERLOG” Software DH-21Figure 4-25. View Data Tabular DisplayThe View Scans Option From “WATERLOG” shown in Figure 4-26 i

Strany 99 - ATERLOG” Software 4-53

DH-21 Using the Graphical “WATERLOG” Software 4-41Figure 4-26. View Scans From WATERLOG OptionWhen first entering this screen, there will be one grap

Strany 100 - ATERLOG” Software DH-21

4-42 Using the Graphical “WATERLOG” Software DH-21Figure 4-27. View Graphical Main Graph

Strany 101 - WATERLOG” software, showing

DH-21 Using the Graphical “WATERLOG” Software 4-43Figure 4-28. View Data DisplayWhen in the “Both Graphs” mode of operation, the main graph will have

Strany 102

Chapter 1IntroductionDH-21 Introduction 1-11.1 Overview of the “WATERLOG”The “WATERLOG” Sensor/Logger, as shown in Figure 1-1, is an innovative water

Strany 103 - ATERLOG” Software 4-57

4-44 Using the Graphical “WATERLOG” Software DH-21Status Boxes:The mouse can be positioned on a graph and the status box will show the information f

Strany 104 - WATERLOG” software

DH-21 Using the Graphical “WATERLOG” Software 4-45Maximize:Both graphs have a box in the upper right corner that when clicked on with the mouse,will

Strany 105 - Guide to Data Analysis

4-46 Using the Graphical “WATERLOG” Software DH-21Figure 4-29. Graph OptionsThe auxiliary graph may be printed by turning on the “Print Auxiliary Gr

Strany 106 - ATERLOG” Logger/Sensor

DH-21 Using the Graphical “WATERLOG” Software 4-47Paper Orientation:The two options here are portrait and landscape. Auto Form Feed:When this option

Strany 107 - Maintenance/Trouble Shooting

4-48 Using the Graphical “WATERLOG” Software DH-21Figure 4-31. Printer OutputFigure 4-31 shows a typical graph of the “WATERLOG” software program.

Strany 108 - ATERLOG” has not moved

DH-21 Using the Graphical “WATERLOG” Software 4-494.4.8 Measure Data OptionThe Measure Data Option will cause the “WATERLOG” to make a measurement as

Strany 109 - Specifications

4-50 Using the Graphical “WATERLOG” Software DH-214.4.9 Zero Sensor OptionThe “WATERLOG” software allows you to force the sensor to use a specific p

Strany 110 - LOGGER SPECIFICATIONS

DH-21 Using the Graphical “WATERLOG” Software 4-514.4.10 Sensor Information OptionYou can see information about your “WATERLOG” sensor as shown in Fi

Strany 111 - DH-21 Specifications A-3

4-52 Using the Graphical “WATERLOG” Software DH-21Figure 4-35. Self-Test OptionFigure 4-36 and Figure 4-37 show the results of the Diagnostics. The

Strany 112 - T BLANK

DH-21 Using the Graphical “WATERLOG” Software 4-53Figure 4-36. Self-Test Diagnostic Display