Xylem Lab 850, starting from serial number 09470315 Uživatelský manuál Strana 52

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 73
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 51
Operation Lab 850
52
ba75890e01 11/2009
4.8.2 Resetting all meter settings
This function resets to the default condition all meter settings including
the calibration values. The corresponding values are given on the
following pages:
Resetting the meter
settings
Note
The measuring system is not calibrated after a reset. Before measuring
recalibrate the meter.
Calibration values page 51
System settings page 43
Measurement settings page 45
1 Switch on the meter with <On/Off>.
The display test appears briefly on the display.
2 During the display test, press <MODE> to open the menu for
the reset of the meter settings.
Init is displayed.
3 Press <><> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.
4 Confirm with <OK>.
The menu is finished.
The meter switches to the measuring mode.
o
n
t
i
n
I
Zobrazit stránku 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 72 73

Komentáře k této Příručce

Žádné komentáře