Multi 3400i Operation
23
Note
At this point, the AutoCal TEC calibration can be terminated with <M>.
This corresponds to a single-point calibration. To do this, the instru-
ment uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the
asymmetry of the electrode.
2 If required, enter the temperature of the first buffer solution,
while keeping the <RUN/ENTER> key depressed, using <>
<>.
3 Immerse the pH electrode in the first buffer solution.
4 Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:
5 When the measured value is stable, Ct2 appears.
/pH
mV
TDS
OU
Tref20
Cal
8
S
1
Sal
%
mg/l
1/
cm
Sal
AutoCal DIN
Lin
Tref25
nLF
RCL
Auto
Store
LoBat
ARng
AR
Time
Baud
Day.Month No.
Ident
Year
mbar
cm
M
°
F
/
K
%
S/
m
cm
8
4
2
t
1
°
C
TP
pH
C
AutoCal TEC
cm
3
7
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
AutoCal DIN
Tref25
Auto
AutoCal DIN
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OO UpHUpH SalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDSTDS
11
SS
mbarmbar
Time
Baud
Day.Month No.
Ident
Year
Time
Baud
Day.Month No.
Ident
Year
0
03
7
pHpH
C
°
C
°
AutoCal TEC
AutoCal TEC
ARAR
mV
88
44
22
TP
TP
/pH
mV
TDS
OU
Tref20
Cal
8
S
1
Sal
%
mg/l
1/
cm
Sal
AutoCal DIN
Lin
Tref25
nLF
RCL
Auto
Store
LoBat
ARng
AR
Time
Baud
Day.Month No.
Ident
Year
mbar
cm
M
°
F
/
K
%
S/
m
cm
8
4
2
t
2
°
C
TP
pH
C
AutoCal TEC
Zobrazit stránku 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 ... 84 85

Komentáře k této Příručce

Žádné komentáře