Multi 3500i Operation
29
4.4.4 Settings for pH and ORP measurements
Overview The following settings are possible for pH and ORP measurements:
z Resolution
z Calibration interval
z Buffers for calibration
z Unit for slope
z Calibration record (display)
Settings The settings are made in the measuring menu of the pH/ORP measure-
ment. To open the settings, activate the relevant measuring window in
the measured value display and press the <MENU/ENTER> key short-
ly. After completing the settings, switch to the measured value display
with <M>.
Menu item Possible
setting
Description
Calibration /
Calibration inter-
val
1 ... 999 d Calibration interval for the pH elec-
trode (in days).
The measuring instrument reminds
you to calibrate regularly by the
flashing sensor symbol in the mea-
suring window.
Calibration /
Calibration type
TEC
NIST/DIN
ConCal
Buffer sets to be used for pH cali-
bration.
For details, see section 4.4.5.
Calibration / Unit
for slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst
slope of
-59.16 mV/pH (100 x determined
slope/Nernst slope).
Calibration /
Calibration
record
- Displays the calibration record of
the last calibration.
Man. tempera-
ture
-20 ...
+130 °C
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Reset - Resets all sensor settings to the
delivery condition (see
section 4.10.1).
High resolution On
Off
Resolution of the pH display:
On = 0.001
Off = 0.01