ProLab 3000 What to do if...
175
ba75645e03 12/2012
CalClock on red
background
15.2 Ion selective measurement
Error message Err1
Obviously incorrect
measured values
Error message, Error
(invalid calibration)
Measurement procedure not
suitable
Follow special procedure
Cause Remedy
Cause Remedy
Calibration interval expired Recalibrate the measuring
system
Cause Remedy
Measuring range exceeded Dilute test sample
Cause Remedy
Electrode not connected Connect the electrode
Cable broken Replace the electrode
Cause Remedy
ISE electrode:
Moisture in the plug Dry plug
Electrode obsolete Replace the electrode
Electrode unsuitable for the
range to be measured
Use a suitable electrode
Socket damp Dry socket
Calibration procedure:
Wrong sequence of standards
for three point calibration
Select correct sequence
Calibration standards do not
have the correct temperature
(max.
± 2 °C temperature differ-
ence)
Adjust the temperature of
the calibration standards